APPENDIX A. SERVICE MENU
Adven
turer ™ Balances Service Manual 87 Ohaus Corporation www.ohaus.com
The sequence of loading the test weights on the platform is very important to insure a correct
linearity calibration. Refer to the diagram below for the correct weight sequence.
Follow the sequence and weights show on the balance display to load and unload weights during
this 3 Point Linearity Calibration.
A.5
Service 5 Point Linearity Calibration
Fig A-8 Service 5 Point Linearity Calibration.
A five point linearization is also rarely required and is the most difficult to perform accurately under
the less than ideal conditions outside the factory. This service calibration requires that accurate
calibration weights be used. Calibration points are done at approximately 0, 25%, 50%, 75% and
100% of maximum capacity.