APPENDIX B. SOFTWARE SERVICE & REPAIR TOOL INSTRUCTIONS
Adven
turer ™ Balances Service Manual 95 Ohaus Corporation www.ohaus.com
B.4
COM Port Configuration.
1. Configure the software Com port configuration to match with the balance and click 'Next'.
Figure B-5. COM Port Configuration.
2. Enter the
balance’s Serial Number and IDNR number.
3. Enter the path to the image file, or click the
Browse
button to locate the file.
4. Click the
Next
button.
5. The software will indicate the download progress. When complete, disconnect the power from
the back of the balance, then re-connect.
6. Perform Service Calibrations (see Appendix A). Test the balance.
Figure B-6. Write Image File.