APPENDIX A. SERVICE MENU
Ohaus Corporation www.ohaus.com 86
Adventurer ™ Balances Service Manual
If the USB Flash drive is not plug in you will see the below message.
Once this selection is made the balance will automatically locate the balance bin file which is store
in the Tum Drive and inserted into the balance front panel USB port, the software upgrade process
will start.
After Software Upgrade re-boot the balance.
A.4
Service 3 Point Linearity Calibration
Fig A-7 Service 3 Point Linearity Calibration.
A three point linearization is rarely required and is difficult to perform accurately under the less
than ideal conditions outside the factory. This service calibration requires that accurate calibration
weights be used. Calibration points are done at approximately 0, 50% and 100% of maximum
capacity.
USB flash drive is not connected.
Current version: 1.00.12
Exit
0%