CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com 22
Adventurer ™ Balances Service Manual
3.2.1 Removing Adventurer Analytical Housing
Turn the balance off and unplug the power cord before you begin.
Preliminary Steps:
For a balance with a Draft Shield:
1. Remove the weighing pan and wind ring and set aside.
Figure 3-1 Analytical Weighing Pan and Wind Ring attach.
Figure 3-2 Analytical Weighing Pan and Wind Ring removed.
2. Remove the EMC plate by removing the two screws holding the EMC plate to the balance top
housing as shown below.
Figure 3-3 EMC Plate Attached Figure 3-4 EMC Plate Removed