OM0302
Rev 2/December 2020
Page 92 of 107
Window Diagnostics
Read inputs all service technician to exercise or check these input function. The inputs may be checked by
altering the input function (i.e. sliding window position) and monitoring the change on the display
Power Window Calibration
The power window can be calibrated by the micro switch position within the right window glide
assembly for major adjustments. For minor adjustments, the window can be stepped up or down to
check for activation points or move into desired access opening height (default window height is 10
inches (254mm). Then the Auto Cal can be used to lock in the window position based on the activation
points of the window high and low microswitches. The Auto Cal will run a sequence, first lowering the
window to the window low position, the raise the window to window high position and back down to the
work access opening height. Smaller changes can be made by adjusting the counts. Low calibration
counts would be when the window is being raised from the low window microswitch and high calibration
counts would be when the window is being lowered from the high window microswitch. Press save to
enter calibration changes.
Summary of Contents for LabGard NU-581-400
Page 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Page 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Page 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Page 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Page 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Page 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Page 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Page 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Page 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Page 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Page 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Page 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...