OM0302
Rev 2/December 2020
Page 81 of 107
Certification Date
This parameter allows the service technician to view and update the current certification date, in addition
to notification time period.
• Filter Maintenance:
This parameter allows the service technician to set the filter status data
used to predict filter life availability. Filter status is based on maximum RPM minus the starting
RPM (entered by technician) then scaled to current RPM to determine filter percentage
availability. Starting RPM data may be entered at any time for service purposes.