OM0302
Rev 2/December 2020
Page 30 of 107
6.1.6.7 system information
This menu item provides the cabinet model number, serial number, and software revision.
6.1.7 Power Window Option
The power window option, if purchased, is factory installed and the display will indicate window up and down
icons on the lower right side of the display. The power window function uses a 24 Vdc motor and internal feedback
encoder to determine position. The power window also incorporates a safety auto-reverse feature to eliminate
any pinch hazard.
Operating the power window is performed by pressing either the UP or DOWN window icon. The travel is
determined by the length of time the icon is pressed. A short press and release will make the window move a
short distance. A long press and release will make the window move to the next standard operating point of either
closed, nominal work access height of 10 inches (254mm) or full open. The power window if desired can be
lowered during the decontamination process for sealing purposes after the armrest is removed.
Summary of Contents for LabGard NU-581-400
Page 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Page 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Page 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Page 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Page 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Page 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Page 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Page 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Page 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Page 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Page 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Page 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...