7.1.2 Auto Decon
The Auto Decon process is intended to be used with any automated cabinet decontamination process (i.e
hydrogen peroxide, chlorine dioxide, etc.) The automated decon process parameters are typically set up
independently of the cabinet. Inlet and outlet ports from the automated decon machine are connected to the
cabinet. The cabinet control system may be set up to run simultaneously aiding the decontaminate circulation
throughout the cabinet during the entire decon cycle.
To validate two decon processes and associated cabinet set up procedure; Nuaire
utilized a Bioquell Clarus II H₂O₂
and a Steris X10.
The following process was performed by the Bioquell Clarus II using the LabGard® ES Model NU-543-600E.
1.
Remove armrest via fasteners.
2.
Remove exhaust shroud and sensor.
3.
Seal top and front openings using either heavy plastic and duct tape or seal plates.
Seal all joints with tape as indicated on drawing BCD-16509.
Note:
Top and front seal plates may be obtained by contacting Nuaire representative or distributor.
4.
Attach inlet and outlet connections to decontamination system along with pressure and/or concentration
sensor if provided.
Summary of Contents for LabGard NU-581-400
Page 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Page 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Page 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Page 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Page 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Page 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Page 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Page 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Page 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Page 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Page 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Page 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...