5. Clean top edge of prefilter with 70% IPA or similar 6a. Rotate both of the prefilter release clamps 90°
cleaner and place wipe in prefilter. Install prefilter
seal over the used prefilter.
6b. Pull up and out prefilter of its position and lay on its
Side to be placed into bag.
7. Place plastic bag over prefilter, gather at one end and tie up with wire tie.