OM0302
Rev 2/December 2020
Page 88 of 107
o
Night Setback Window Interlock
This parameter allows for the selection of the night setback function to be initiated upon a sliding
window closure. Once the sliding window is closed, the LED light will be disabled, and the airflow will
be reduced to a percentage of the normal setpoint.
NOTE:
Blower window interlock must be turned off to allow the night setback function to operate.
Auxiliary Features:
This parameter allows the access to turn on or off the auxiliary relay function, accessory outlet and power
window.