74
Change Certificate Language
The language will change for all New and Archived Calibration Certificates.
Select
Options
then select
Certificate Language
then select language.
TIP: If a certificate was open before the language change, close the certificate and re-open.
Reset Test Data Read Index
This should only be selected if you are experiencing loss of data transfer between the T-Box XL™ and
TDMS using either USB synchronization or a USB stick.
Every piece of test data T-Box XL™ stores has a unique ID in increasing values. So that only the latest data is
transferred, TDMS only requests test values with an ID greater than the one it last read for that T-Box XL™
serial number. When you do a reset it sets this read ID to zero so all
test data is transferred. However it also
stores the test ID and won't duplicate data it already has stored.
TIP: If you have already synchronized T-Box XL™ to TDMS and select Reset Test Data Read Index,
the next time that you synchronize with TDMS, some data may be duplicated.
Summary of Contents for T-Box 43236
Page 2: ......
Page 47: ...45 4 Select Yes No or Cancel...