62
If the measurement device and transducers uncertainty values are too high, the following message will
appear:-
Check the measurement device and transducer uncertainty values are correct.
The measurement systems maximum relative error can be no greater than ¼ of the tools maximum
permissible relative deviation.
The measurement systems maximum relative error is calculated by taking the transducers expanded
uncertainty value from it uncertainty interval value.
The tools maximum permissible relative deviation uses the ‘Expected measurement error’ value under the
‘Test’ tab of the tool template. If this value is blank or is greater than the default maximum permissible
relative deviation defined in ISO6789-1:2017 (section 5.1.5), the ‘Upper Limit’ and ‘Lower Limit’ values under
the same tab will be used.
If the tool was conformance tested below 18°C or above 28°C the following message will appear:-
Check the correct temperature was entered.
If the measurement system and temperature comply with ISO 6789-1:2017, the declaration of conformance
will be generated:-