12. Select the ‘Test’ tab
13. Select
Modes Available
by ticking / un-ticking the boxes.
Only the modes ticked (selected) will be available when the Tool Template is used on a T-Box XL™.
14. Enter
Number of Tests
. This is the number of readings taken at each Target value.
This is greyed out for ‘ISO 6789’ Tool Types.
15. Enter
Upper Limit
. This is the Upper Calibration Limit and is a percentage of reading at each Set Point.
This is greyed out for ‘ISO 6789’ Tool Types.
16. Enter
Lower
Limit
. This is the Lower Calibration Limit and is a percentage of reading at each Set Point.
This is greyed out for ‘ISO 6789’ Tool Types.
17. Click each
Set Point (%)
to change if required.
Only the number of values for the tool type will be available.
The value can be up to 3 decimal places i.e. 26.087 for ISO 6789 & Other type Tool Templates.
These must be entered incrementing from the smallest to the largest.
Click
Change to Values
/
Change to %
to enter set points in torque units or %.
Click
Default
to set values to 20%, 60% and 100% if required.
18. Enter
Expected measurement error
. This is the expected maximum percentage error of reading for
this model of tool.
This is greyed out for ‘Other’ Tool Type.
19. Enter
Expected relative uncertainty interval
. This is the expected maximum relative uncertainty
interval, expressed as a percentage value, for this model of tool.
This is greyed out for ‘Other’ Tool Type.
Note: If this value is not yet known, leave it blank until a value is established from calibrating more
specimens of this model of tool.