57
Technical Information
System Specifications
Power Requirements:
Power Supply: 100 – 240 VAC 50 – 60 Hz
Power Consumption: 3.6 W with battery operation
9 W with power supply
Input: 12 VDC, 720 mA
Internal Battery:
Type: Lithium Ion (Li-Ion) internal battery, non-field replaceable,
rechargeable
Battery Capacity: Approximately 5 hours
Charging Time: Approximately 9 hours
Dimensions:
200 x 135 x 50 mm (7.9 x 5.3 x 2 in.)
Weight:
865 grams (1.9 pounds)
Temperature:
Operating: 0 °C to 40 °C (32 °F to 104 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
Time (from storage) for monitor to be ready
for its intended use:
10 minutes to warm from -40 °C to ambient temperature of
20 °C
19 minutes to cool from 70 °C to ambient temperature of
20 °C
Applied part temperature will not exceed 41°C as measured during a controlled environment test.
Humidity:
Operating: 15% to 93% noncondensing
Storage/Transportation: Up to 93% noncondensing
Altitude:
Operating: Up to 2,740 meters (9,000 feet)
Atmospheric Pressure: 720 to 1060 hPa (540 to 795 mmHg)
Pump:
Pump Flow: 75 ml/min
Flow Accuracy: ±15 ml/min
High and Medium Priority Alarms (at 1 m):
Sound Pressure Level: 74.6 to 75.5 dB
Classification per ANSI/AAMI ES60601-1 and CSA C22.2 No. 60601.2:
Type of Protection: Internally powered class II (with power supply)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection: IP22