27
Using the RespSense II Monitor
Minimum Alarm Volume
This setting controls the monitor’s available audible alarm volume levels.
The default minimum alarm volume is Medium.
Set the Minimum Alarm Volume
1. Access the Responsible Organizations Settings screen.
2. In the Minimum Alarm Volume field, press one of the options. Options include Off, Medium
(default), and Maximum. A check mark displays to the left of the selection.
• Off – all options are available when monitoring, including shutting audible alarms off.
• Medium – both medium and maximum alarm volumes are available when monitoring.
• Maximum – only the maximum alarm volume is available when monitoring.
Default Alarm Volume
The default alarm volume is Maximum.
Set the Default Alarm Volume
1. Access the Responsible Organizations Settings screen.
2. In the Default Alarm Volume field, press one of the volume options. Options include
Medium and Maximum (default). A check mark displays to the left of the selection.
Alarm Limits – EtCO
2
, RR
The alarm limit settings allow the Responsible Organization to restrict upper and lower alarm
limits.
Use the up/down arrows to adjust the high and low alarm limits to the desired values.
Set the Responsible Organization Alarm Limits
1. Access the Responsible Organizations Settings screen.
2. In the EtCO
2
field, use the up/down arrows to adjust the high and low alarm limits to the
desired values.
3. In the RR (respiration rate) field, use the up/down arrows to adjust the high and low alarm
limits to the desired values.