15
Display Screens
Display Screens
The following section describes the icons and their functions on the select alarm group and
operating screens.
Select Alarm Group Screen
NOTE:
This screen does not display if the monitor has locked alarm limits (see
section).
The select alarm group screen (figure 3) displays after the start-up screen. On this screen, the
operator may select alarm limits for a patient under 30 kg, over 30 kg, or select the last used alarm
settings.
If a selection is not made within approximately 8 seconds, the monitor defaults to the over 30 kg
alarm limits.
Figure 3. Select Alarm Group Screen
1
2
3