16
Display Screens
Table 3. Select Alarm Group Screen – Display Descriptions
No.
Symbol
Description
1
Alarm Limits
≤
30kg / 66lbs
Pressing this icon selects the default alarm limits for patients
weighing 30 kg (66 lbs) or less.
•
If the Responsible Organization Settings have not been
set, the alarm limits will be the factory default settings
(see table 5).
•
If the Responsible Organization Settings are set, the
alarm limits may be restricted by the Responsible
Organization alarm limits.
2
Alarm Limits >30kg / 66lbs
Pressing this icon selects the default alarm limits for patients
weighing more than 30 kg (66 lbs).
•
If the Responsible Organization Settings have not been
set, the alarm limits will be the factory default settings
(see table 5).
•
If the Responsible Organization Settings are set, the
alarm limits may be restricted by the Responsible
Organization alarm limits.
3
Last Used Alarm Setting
Pressing this button selects the last used alarm settings and
alarm volume.