Maintenance and Inspection
Recommended Inspections and Functional Check
1. Before each use, verify the equipment is clean and in optimal operating condition. See
2. Verify the single-use, disposable sample line or cannula is free of bends and kinks for
optimal performance.
3. Verify the single-use, disposable moisture trap and filter are in position.
4. Visually examine the accessories for defects prior to use.
5. Press
On/Standby
to turn the monitor on.
6. Verify battery capacity. If the battery indicator is yellow or red, connect the monitor to the
power supply and a power outlet.
NOTE:
To ensure the monitor is always ready to use, always connect the monitor to an outlet
whenever it is not in use.
7. Verify all parameters display correctly and, if allowed, adjust any alarm limits according to
the patient.
8. Verify alarm function/status by simulating alarm situations for all parameters.
9. Visually verify the zero-point of the CO
2
graph is not elevated.
CAUTION:
After exposing the monitor to an environment outside of normal room
temperature/humidity conditions, always replace the moisture trap and filter before
each use.
WARNING: If the monitor fails to respond as described, discontinue use and contact
Nonin Technical Service.
CAUTION:
RespSense II should only be operated by trained licensed practitioners.
WARNING: Never allow liquids to enter into or to be spilled onto the monitor. If liquid
has penetrated into the monitor it must be checked by Nonin Technical Service.
CAUTION:
Be careful not to drop the monitor on the floor or strike it against hard
surfaces. If such an incident happens, do not use the monitor until a functional test
has been carried out.
CAUTION:
This device is a precision electronic instrument and must be repaired by
qualified technical professionals. Field repair of the device is not possible. Do not
attempt to open the case or repair the electronics. Opening the case may damage
the device and void the warranty.
!
!
!
!