Using the RespSense II Monitor
Responsible Organization Settings Screen
This screen allows the operator to establish institution default high and low limits for EtCO
2
and
respiration rate.
This screen is accessed through the Configuration Menu (see
NOTES:
•
The operator is required to enter a new 4-digit PIN the first time the Responsible
Organization Settings screen is accessed or after restoring factory defaults.
•
The Responsible Organization Settings are populated with factory default settings until
changed by the organization.
•
Once a PIN is saved, the alarm limits become restricted to the high and low limits on the
Responsible Organization Settings screen.
1. On the Configuration Menu, press
Responsible Organization Settings
.
2. Enter the responsible organization PIN and press
OK
. Responsible Organization Settings
screen displays (figure 9).
3. From this screen, the operator may:
• Set the minimum alarm volume
• Set the default alarm volume
• Set the responsible organization’s alarm limits: EtCO
2
and respiration rate (RR)
• Reset the device to the factory defaults
• Change the 4-digit PIN
4. Press
Save
when all changes to the Responsible Organization Settings are complete. To
exit without saving changes, press
Cancel
.
Figure 9. Responsible Organization Settings Screen