Nikon N Series/K Series Total Station Instruction Manual
You can delete poor observations and recalculate if necessary. You can also select the
BS point.
1.
To start the resection, press
[2]
or select
Resection
in the Stn Setup menu.
2.
Enter the point name for the first observation
point (PT1).
3.
Enter the target height and press
[ENT]
.
4.
Sight PT1 and press
[MSR1]
or
[MSR2]
.
To use the
F2
softkey for F1/F2 measurements,
Advanced feature: Measuring F1 and F2,
5.
To proceed to the next point, press
[ENT]
.
6.
Enter the second point (PT2) and its height of
target.
7.
Measure to PT2 and press
[ENT]
.
When the instrument has enough data, it
calculates the station (STN) coordinates.
If the angle between known point 1 and known point 2 (measured from the station point) is extremely acute or
extremely oblique, the resulting solution will be less reliable geometrically. For geometric reliability, select known
point locations (or station point locations) that are widely spaced.
The
F2
softkey is not available on the N 5" and K 5" single face model.
Summary of Contents for K Series
Page 2: ......
Page 12: ...Safety xii Nikon N Series K Series Total Station Instruction Manual ...
Page 18: ...Contents xviii Nikon N Series K Series Total Station Instruction Manual ...
Page 70: ...3 Getting Started 52 Nikon N Series K Series Total Station Instruction Manual ...
Page 162: ...5 Menu Key 144 Nikon N Series K Series Total Station Instruction Manual ...
Page 168: ...6 Checking and Adjustment 150 Nikon N Series K Series Total Station Instruction Manual ...
Page 172: ...7 System Diagrams 154 Nikon N Series K Series Total Station Instruction Manual ...
Page 188: ...8 Communications 170 Nikon N Series K Series Total Station Instruction Manual ...