6
Checking and Adjustment
148
Nikon N Series/K Series Total Station Instruction Manual
Adjusting
To enter the calibration screen, press
[MENU]
and
[7]
.
1.
The N Series/K Series has vertical axis
adjustment only. Take an F1 measurement to a
target on the horizon. Press
[ENT]
.
The vertical angle is shown in the
V0 dir= Horiz setting.
When you have taken the measurement, the
message on the bottom line changes from
DO
NOT TOUCH!
to
Turn to F2
.
2.
Take an F2 measurement to the same target.
Press
[ENT]
.
When the observation on F2 is completed, four
parameters are displayed.
3.
Do one of the following:
–
To return to the first observation screen,
press
[ESC]
or the
Redo
softkey.
–
To set parameters on the instrument, press
[ENT]
or the
OK
softkey.
VA1
Face-1 vertical angle (tilt-off value)
HA1
Face-1 horizontal angle (tilt-off value)
Tilt
Face-1 Y axis tilt value
VA2
Face-2 vertical angle (tilt-off value)
HA2
Face-2 horizontal angle (tilt-off value)
Tilt
Face-2 Y axis tilt value
If ACV, ACH,Tilt is out of range, OVER appears. Press any key to return to the
first observation screen.
Summary of Contents for K Series
Page 2: ......
Page 12: ...Safety xii Nikon N Series K Series Total Station Instruction Manual ...
Page 18: ...Contents xviii Nikon N Series K Series Total Station Instruction Manual ...
Page 70: ...3 Getting Started 52 Nikon N Series K Series Total Station Instruction Manual ...
Page 162: ...5 Menu Key 144 Nikon N Series K Series Total Station Instruction Manual ...
Page 168: ...6 Checking and Adjustment 150 Nikon N Series K Series Total Station Instruction Manual ...
Page 172: ...7 System Diagrams 154 Nikon N Series K Series Total Station Instruction Manual ...
Page 188: ...8 Communications 170 Nikon N Series K Series Total Station Instruction Manual ...