INC
JAA79551-R.3690.A
- A 63 ・ AF-S VR 70-300/4.5-5.6G -
⑱
When the shutter of the camera is released, slide the tripod to the back by 10
±
0.1 cm and make a
remeasurement.
⑲
Again, slide the tripod to the back by 10
±
0.1 cm and make a remeasurement.
Repeat this operation four more times, totalling in seven measurements.
(The total sliding distance is 60 cm.)
Note 1: When the below warning is given, there may be some defects in the brightness of the viewers and/
or parallelism of the chart and camera, etc. So correct the above and make a remeasurement.
Note 2: When the below warning is given, recheck that the Quality mode of the camera is set to
RAW.