97
DEVICE SETTINGS AND MAINTENANCE: Cleaning
4
Check if the window is cleaned using a penlight. If soiled areas remain, clean the
window again with new cleaning paper.
4.7.2
Cleaning the printer
After repeated usage, the paper slot of the auto cutter of the printer may become soiled with powdery
paper residue. If paper residue settles, malfunction of the auto cutter may result.
1
Open the printer cover and remove the printer
paper roll.
2
Apply the nozzle of a vacuum cleaner to the auto
cutter to remove paper residue.
Never blow off paper residue with a blower. If paper resi-
due settles on the internal working structure, malfunction
may result.
3
Supply the printer paper as it was.
• When the “61. WINDOW CHECK” parameter is set to “YES” or “DAY”, the measuring window is
checked for cleanliness at device start-up.