OPERATING PROCEDURE: Operation when Peripheral Devices are Connected
3.13.2 Operating procedure
●
Importing data from a lensmeter (LM) (RS-232C connection)
1)
After measurement by a lensmeter, press the print button.
●
Exporting data to the RT (or PC) (RS-232C connection)
After measurement, press the print
button.
1)
Transmit data to the RT (or PC).
When the device is connected to the RT, it receives data number (ID number) from the RT.
When the device is connected to a PC, it does not receive data number (ID number).
2)
The measured data is printed.
When the device is connected to the RT, the data number (ID number) is also printed.
●
Exporting data to the PC (network connection [LAN])
After measurement, press the print
button.
The measured data is printed and data is transmitted to the PC.
●
Reading the patient ID with the barcode reader or magnetic card reader
ARK-1s parameter setting
NIDEK LM parameter setting
73. BAUD-RATE = 9600
Printer = AR print
74. BIT LENGTH = 8
RS-232C = NIDEK
76. LM DATA PRINT = YES
Baud rate = 9600
Parity = Odd
Data bits = 8
Stop bits = 1
• Although patient ID can be read before or after measurement, read it before printing the measured
results.
If patient ID is read after measured data has been printed and is still displayed, the device considers the
displayed data to be that of a former patient and erases it automatically.
• The device considers the latest patient ID read before printing to be the patient ID of the printed data.
If an incorrect patient ID has been read, read the correct ID again.
• A beep sounds when the device is turned on with the barcode scanner or magnetic card reader con-
nected.