OPERATING PROCEDURE: Manual Measurement
5
Press the right
button or left
button to align
the guide lines on the pupil of the patient’s eye.
The guide line to be aligned is displayed in pink. Press the
down
button to change the selected guide line.
6
Press the start button to confirm the measurement.
A PS value (0.1 mm increments) is displayed in the lower part
of the screen.
7
Measure the other eye in the same manner.
Switching the eye to be measured displays the Step 1 screen display.
8
Press the return
button to exit from PS mea-
surement mode.
The screen returns to the measurement screen.
The PS-measured value is displayed on the screen indicating
the completion of PS measurement.
To perform PD (Pupillary Distance) measurement at the same
time, press the CS/PS/PD
button to display the PD
measurement screen.
ᵏ
383,/6,=(
/$0321
ᵏ
383,/6,=(
/$0321
PP
• When the “5. AI MODE” parameter is set to “YES” and the “31. PRINT” parameter is set to “AUTO”,
PS measurement must be performed before AR and/or KM measurement to print the data together
with AR- and/or KM-measured data.