SPECIFICATIONS AND TECHNICAL INFORMATION: EMC (Electromagnetic Compatibility)
5.3
EMC (Electromagnetic Compatibility)
The device complies with the International Electrotechnical Commission standards (IEC 60601-1-2:
2007) for electromagnetic compatibility as listed in the tables below. Follow the guidance in the tables
for use of the device in an electromagnetic environment.
O
Specified cable
O
Essential performance
Objective refraction function
Keratometry measurement function
WARNING
•
Use the specified accessories, optional accessories, and cables.
Otherwise increased emissions or decreased immunity of the device may result.
•
Do not use the device near, on, or under other electronic equipment.
CAUTION
•
The device needs special precautions regarding EMC.
The device needs to be installed and used in accordance with the EMC information provided in this
manual.
•
Portable and mobile RF communications equipment can affect the device.
Part name
Connector
Shielded
Cable Shielded
Ferrite Core
Length (m)
Power cord (100 V)
No
No
No
2.5
Power cord (230 / 240 V)
No
No
No
2.5
LAN cable
Yes
No
No
5.0
USB card reader cable
Yes
Yes
No
2.0
RS-232 cable
Yes
Yes
No
5.0