DEVICE SETTINGS AND MAINTENANCE: Device Parameter Settings
●
FUNCTION (Various functions)
●
COMMUNICATION (RS-232C communication function)
Parameter option
Setting contents
61. WINDOW CHECK
(Measuring window
check function)
YES / NO / DAY
Sets whether to automatically check the measuring window for soiling.
DAY
: The measuring window is checked at the first startup of the day.
YES
: The measuring window is checked every startup.
NO
: The measuring window is not checked.
62. TRACKING SW
(Tracking button)
NORMAL / ALL
Selects the type of the auto tracking function (3D/2D/OFF) and auto shot function
(ON/OFF) by pressing the auto
button.
NORMAL
: Select from “3D/ON”, “3D/OFF”, or “OFF/OFF”.
ALL
: Select from “3D/ON”, “3D/OFF”, “2D/ON”, “2D/OFF”, “OFF/ON”, or “OFF/
OFF”.
63. SLEEP
(Sleep time)
5MIN / 10MIN / 15MIN / NO
Selects the time to enter sleep mode when the device has not been used.
64. BEEP
(Beep sound)
LOW / HIGH / NO
Selects the pitch of the beep sound (electronic sound) produced during measure-
ment or such.
65. LCD BRIGHTNESS
(Brightness of LCD)
LOW / MEDIUM / HIGH
Sets the brightness of the LCD.
66. ICON OFF
(Icon display)
YES / NO
Sets whether to display function icons on the measurement screen.
YES
: Function icons are not displayed during measurement. Function icons are
automatically displayed when measurement is complete. Function icons are dis-
played after a specified period of time or function button is pressed.
NO
: Function icons are always displayed.
67. MAN FOCUS DISP
(Manual focusing indica-
tor display)
YES / NO
Sets whether to display the focusing indicator in manual mode (auto tracking OFF +
auto shot OFF).
68. AUTO PD
(Automatic PD measure-
ment)
YES / NO
Sets whether to measure the Pupillary Distance (PD) automatically for AR mea-
surement.
69. AUTO PS
(Automatic PS measure-
ment)
YES / NO
Sets whether to measure the Pupil Size (PS) automatically for AR measurement.
70. AUTO CS
(Automatic CS measure-
ment)
YES / NO
Sets whether to measure the Corneal Size (CS) automatically for AR measurement.
Parameter option
Setting contents
71. I/F MODE
(Communication mode)
NIDEK / NIDEK2 / NCP10
Communication setting of connected device
NIDEK
: Communication with a NIDEK-brand product
NIDEK2
: Communication with a NIDEK-brand product (extended timeout period)
NCP10
: Communication with a NCP10-compliant device
NIDEK2 is a communication mode whose timeout period is extended over that of
the NIDEK setting. Set the mode to NIDEK2 when timeout occurs for a communi-
cation environment using the NIDEK setting.