NEOMERIS CONTROL DES
Installation
Build:20211025-094624-9873335NSt
All information and technical data are subject to
change
Page 18
5
Installation
5.1
Preparing for installation
For setup and installation, please refer to the existing plans and draw-
ings as in chapter 4 "Specifications."
We recommend interconnecting alerts with the control room. Ignoring
or failing to acknowledge the fault over a longer period of time may lead
to severe damage to the system or even a complete production down-
time.
5.1.1
Setup location
Ensure that the setup location is frost-free and sufficiently ventilated,
well lit and clean.
Condensation in the controller may ruin it!
5.1.2
Setup area
Ensure that the wall space is flat. Place the controller in a well-lit and
easily accessible place so that operating and display elements are eas-
ily visible and accessible.
Do not place the controller on surfaces with high vibration (e.g., cooling
tower walls). This may damage the electronic system!
5.1.3
Power supply
Ideally, connect the controller to a power supply with a 6A (10A max.)
fuse. We recommend protecting the in-house mains installation from
lightning. In any case, it is necessary to protect the power supply from
lightning.
5.1.4
Cabling
It is not admissible to use the same cable for measuring lines (or
control signals) and mains supply, even if there is a sufficient
number of unused wires!
Hazard of damage or disruptions from electro-magnetic fields!
§
Installing the device or the connection lines parallel to the mains
cable or near strong electro-magnetic fields may damage the de-
vice or cause disruptions during measurement.
!
!
WARNING
!
!
CAUTION
!
!
CAUTION
i
NOTE
i
NOTE
i
NOTE