Managing the Data Port
86
10030881 Rev. B 12-2008
Service Manual
The RS-232 hardware interface cable has a DB-15 connector for the OxiMax N-600x
pulse oximeter and the applicable connector for the Philips monitor. For this
interface, Nellcor recommends Nellcor cable part number 902256.
A blank screen on the Philips monitor indicates corrupt data. The Philips monitor
detects corrupt data in less than 100 milliseconds. In the Philips mode of operation,
the OxiMax N-600x pulse oximeter automatically sets the interface baud rate to
19200 bits per second (bps).
Managing the Data Port
Data Printout Information
Printing Oximeter Trend Data
Trend information (oximeter and in-sensor event history) can be sent to a personal
computer or to a serial printer.
Note:
The protocol settings must be set to ASCII MODE for printing text data.
To print trend data
1. With the oximeter in the normal monitoring mode, connect the serial printer to
the oximeter’s data port connector, using Nellcor printer cable part number
036341.
Figure 41.
Data Port Location
2. Turn on the printer.
3. Press the SETUP softkey on the monitor, and then press the NEXT softkey twice.
4. Press the COMM softkey.
5. Set the BAUD rate to the appropriate number using the ADJUST UP or ADUST
DOWN key.
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......