Using Oximeter Softkey Menus
Service Manual
10030881 Rev. B 12-2008
67
TREND Menu
Choose the type of trend data to view by selecting trend data from the oximeter
(MONITR) or trend data from the pulse oximetry sensor (SENSOR) in the Trend menu.
1.
MONITR Menu
—Isolate oxygenation (SpO
2
) or pulse (PULSE) trend data or view
them both together (DUAL) for a specified length of time. Establish trend data
parameters for specific time segments and set minimum and maximum trend
values.
a.
VIEW Menu
—Isolate oxygenation (SpO
2
) or pulse (PULSE) trend data or view them
both together (DUAL). Access the HIST and Amplitude submenus by selecting the
NEXT Option.
-
DUAL softkey
—View oxygenation
(
SpO
2
) and pulse (BPM) trend data together
for a specified length of time.
-
SpO
2
softkey
—Isolate oxygenation
(
SpO
2
) trend data for a specified length of
time.
-
PULSE softkey
—Isolate pulse (PULSE) trend data for a specified length of time.
-
HIST softkey
—Choose to delete or keep current trend data history. To delete,
select YES. To keep, select NO and return to the HIST menu, go BACK, or EXIT to
the Main Menu.
-
AMP softkey
—Set the trend data view to “Pulse Amplitude Units” (PAU) using
the AMP (amplitude) Menu. This is an arbitrary unit of measure gauging the
distance from the peak to the valley of a pulse waveform. The trend data when
set to PAU can register amplitudes up to 200 PAU.
b.
ZOOM Menu
—Access to the TIME option, SCALE option, AUTO option, and BACK
option if the trend view is set in DUAL, SpO
2
, or PULSE mode.
Note:
The ZOOM menu will not allow access to the options listed if it is in AMP mode. SCALE and
AUTO options do not appear as options without trend data history to review.
-
TIME softkey
—Select from hours to minutes to seconds for viewing a specific
time segment of trend data. Cycle through the options by pressing the TIME
softkey. The options proceed in the following order: 48h, 36h, 24h, 12h, 8h, 4h,
2h, 1h, 30m, 15m, 40s, or 20s.
-
SCALE softkey
—Select the maximum and minimum values for the SAT or
PULSE trend graph. Default SAT trend values are 10 to 100. Default PULSE trend
values are 5 to 250. Cycle through the options by pressing the Select softkey.
The options proceed in the following order: ±5, ±10, ±15, ±20, ±25, ±30, ±35,
±40 and ±50 units variance.
-
AUTO softkey
—Obtain the maximum and minimum rounded values based on
all graphed trend data.
-
DELETE softkey
—Choose to delete or keep current trend data. To delete,
select YES. To keep, select NO and return to the previous menu.
To Purchase, Visit
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......