Troubleshooting
142
10030881 Rev. B 12-2008
Service Manual
1.
Prompts
—Prompts require a response. For example, the “SAVE DEFAULTS
?”
prompt requires a user response of YES or NO.
Figure 118.
Save Defaults Prompt Screen
2.
Error messages
—Error messages provide information. The
SENSOR
DISCONNECTED
error message leaves any action to the discretion of the user.
Advisory messages appear as centered text at the bottom of the screen.
Figure 119.
Sensor Disconnected Message Screen
Table 24.
Common Prompts and Error Messages
Message
Time-
out in
secs
Exit on
Alarm?
Exit on
Alarm
Silence?
Displayed
Resolution
ADJUST
CONTRAST
UP/DOWN
10
Yes
Yes
When pressing and
holding the HELP/
CONTRAST key.
To adjust the contrast, use the
ADJUST UP to increase the
contrast or ADJUST DOWN
key to decrease contrast.
CLOCK
SETTING LOST
None
No
No
If the oximeter
detects that the real-
time clock has
stopped running and
both battery and AC
power are lost.
Power-cycle the oximeter.
If the error remains, recharge
or replace the battery.
DATA IN
SENSOR
5
No
Yes
When a sensor
containing data is
connected to the
oximeter.
Clears on time-out, sensor
disconnect, or pressing the
ALARM SILENCE key,
whichever comes first.
DATA TYPE:
SpO
2
5
No
Yes
When a blank event
sensor is connected
to oximeter with the
event data type set to
SpO
2
.
Clears on time-out, sensor
disconnect, or pressing the
ALARM SILENCE key,
whichever comes first.
To Purchase, Visit
Summary of Contents for OXIMAX N-600X Series
Page 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Page 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Page 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Page 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Page 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Page 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Page 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Page 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Page 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Page 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Page 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Page 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Page 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Page 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Page 209: ......