BenchMike Pro Instruction Handbook: Specifications
8-3
8.2.2
BenchMike Pro Model 2025 Specifications
1
Specification
Model 2025
Measurement Range
0.100 to 25.4 mm (0.004 to 1.0 in.)
Repeatability
2
±0.25 µm (±0.000010 in.)
Linearity
3
±0.9 µm (±0.000036 in.)
Measurement area depth of field
±0.75 x 25 mm (±0.030 x 1.0 in.)
Temperature coefficient
±0.008 µm/mm°C
(
±0.000004 in./in.°F)
Laser beam spot size
100 µm (0.004 in.)
Laser beam velocity
50 m/sec. (2,000 inch/sec.)
8.2.3
BenchMike Pro Model 2050 Specifications
1
Specification
Model 2050
Measurement Range
0.254 to 50 mm (0.010 to 2.0 in.)
Repeatability
2
±0.5 µm (±0.000020 in.)
Linearity
3
±1.5 µm (±0.000060 in.)
Measurement area depth of field
±1.5 x 50 mm (±0.060 x 2.0 in.)
Temperature coefficient
±0.008 µm/mm°C
(
±0.000004 in./in.°F)
Laser beam spot size
250 µm (0.010 in.)
Laser beam velocity
100 m/sec. (4,000 inch/sec.)
Notes:
1
Specifications valid after a warm-up period of four hours in a stable temperature environment.
2
Specified repeatability is the maximum deviation (±) from the mean value of 90 consecutive readings, with 2-
second averaging. (A 2-second averaging reading is the average of 200 single scans.) Test conditions include
using a gauge pin with a low coefficient of expansion, without removing the gauge pin and with minimal air flow
around the working area of the BenchMike Pro.
3
Specified linearity confirmed using standard factory calibration @ 68 °F at 50% relative humidity. Testing condi-
tions include a measured part composed of material with low coefficient of expansion, with minimal air flow