6-20
BenchMike Pro Instruction Handbook: Remote Setup
valid data unless the appropriate measurements
have been taken. Examples:
MEAS.LISTVARS?
MEAS.RES
Request or specify the measurement resolution
(number of significant digits). Options are numbers 0
through 7. Examples:
MEAS.RES:5
MEAS.RES?
MEAS.ROUND
Request or specify the measurement rounding. Op-
tions are: NEAREST, EVEN, 0/5, and TRUNCATE. Ex-
amples:
MEAS.ROUND:NONE
MEAS.ROUND?
MEAS.SENDDATA.NOW
Request the last measurement be transmitted. Use
CONF.SER.FORMAT commands to change or view the
format of this data.
MEAS.SENDDATA.NOW:
MEAS.START.NOW
This command informs the BenchMike Pro to begin
sending measurements from the scanner immediate-
ly. Example:
MEAS.START.NOW:
MEAS.STOP.NOW
This command informs the BenchMike Pro to stop
sending measurements from the scanner immediate-
ly. Example:
MEAS.STOP.NOW:
MEAS.UNITS
Request or specify the measurement units. Options
are: INCH (or IN), MILS (or MIL), MICROINCH (or UI),
CM, MM, and MICRON (or UM). Examples:
MEAS.UNITS:MM
MEAS.UNITS?