National Instruments Corporation
35
NI 5622 Calibration Procedure
4.
Configure the common vertical properties using the niScope Configure
Vertical VI.
5.
Configure the horizontal properties using the niScope Configure
Horizontal Timing VI.
6.
Configure the analog dither using the niScope Property Node.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVertical
with
the following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
range
:
1.0
offset
:
0.0
coupling
:
NISCOPE_VAL_AC
probeAttenuation
:
1.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureHorizontal
Timing
with the following parameters:
vi
: The instrument handle from
niScope_init
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
minSampleRate
:
150,000,000
refPosition
:
0.0
minNumPts
:
31,000,000
LabVIEW VI
C/C++ Function Call
Call
niScope_ SetAttributeViBoolean
with
the following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
attributeId
:
NISCOPE_ATTR_DITHER_ENABLED
value
:
NISCOPE_VAL_TRUE