NI 5622 Calibration Procedure
44
ni.com
7.
Commit all the parameter settings to hardware using the niScope
Commit VI.
8.
Initiate a waveform acquisition using the niScope Initiate
Acquisition VI.
9.
Acquire a waveform using the niScope Fetch (poly) VI. Select the
WDT instance of the VI. Use the default value for the
timestamp Type
parameter.
LabVIEW VI
C/C++ Function Call
Call
niScope_Commit
with the
following parameter:
vi
: The instrument handle from
niScope_init
LabVIEW VI
C/C++ Function Call
Call
niScope_InitiateAcquisition
with the following parameter:
vi
: The instrument handle from
niScope_init
LabVIEW VI
C/C++ Function Call
Call
niScope_Fetch
with the following
parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
timeout
:
5.0
numSamples
:
–1