NI 5622 Calibration Procedure
42
ni.com
Average Noise Density
Complete the following steps to verify the average noise density of the
NI 5622.
1.
Connect the 50 ohm terminator directly to channel 0 of the digitizer.
2.
Open a session and obtain a session handle using the niScope
Initialize VI.
3.
Configure the input impedance and the maximum input frequency
using the niScope Configure Chan Characteristics VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the following
parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the following
parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
inputImpedance
:
NISCOPE_VAL_50_OHM
maxInputFrequency
:
250,000,000