National Instruments Corporation
43
NI 5622 Calibration Procedure
4.
Configure the common vertical properties using the niScope Configure
Vertical VI.
5.
Configure the horizontal properties using the niScope Configure
Horizontal Timing VI.
6.
Configure the bandpass filter using the niScope Property Node.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVertical
with the following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
range
: The
Range
value in Table 8for the
current iteration
offset
:
0.0
coupling
:
NISCOPE_VAL_AC
probeAttenuation
:
1.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureHorizontal
Timing
with the following parameters:
vi
: The instrument handle from
niScope_init
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
minSampleRate
:
150,000,000
refPosition
:
50.0
minNumPts
:
524,288
LabVIEW VI
C/C++ Function Call
Call
niScope_SetAttributeViBoolean
with
the following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
attributeId
:
NISCOPE_ATTR_BANDPASS_FILTER_ENABLED
value
: The
Bandpass Filter
value listed in
Table 8 for the current iteration