National Instruments Corporation
47
NI 5622 Calibration Procedure
Figure 5.
NI 56222 Connections for External Adjustment
2.
Obtain a calibration session handle using the niScope Cal Start VI.
3.
Configure the signal generator to generate a sine wave with
Input Frequency
and
Input Voltage
for the current iteration in Table 9.
4.
Wait the amount of time the manufacturer recommends for the output
to settle.
1
R&S NRP-Z91 Power Meter
2
Weinschel WA 1507R Splitter
3
To Signal Generator
LabVIEW VI
C/C++ Function Call
Call
niScope_CalStart
with the
following parameters:
resourceName
: The device name assigned
by MAX
password
:
The password required to open
an external calibration session. If this
password has not been changed since
manufacturing, the password is
“NI”
NI PXIe-5622
16-Bit IF Digitizer