Test result issues (cont.)
EMTF-03 -
Unable to output a test result because the system detected
readings in the cartridge during analysis that were abnormal so a
result could not be calculated.
• Load the same sample into a new cartridge and load the cartridge into
the system. If the problem is repeated, try a sample from a different source
(different patient). If the problem still continues, contact customer support.
EMTF-04 -
The lateral flow of the sample within the cartridge is
incomplete.
•
Load the sample into a new cartridge and restart the test. If the same error
message is displayed, centrifuge the sample to remove particulate matter and
fibrin clots and retry with a new cartridge. If the problem persists, try a different
sample. If the problem still continues, contact customer support.
Trouble-
shooting
Maintenance