FREND
TM
System © 2012 NanoEnTek Inc.
47
Barcode issues
EMBR-03 -
The lot number entered for the cartridge does not match
the lot number on the code chip.
• Carefully enter the lot number of the cartridge again. If the same
message is repeated, check the lot number on the installed code
chip. If code chip and cartridge lots do not match, change either code
chip or cartridge so they are the same lot number and retry loading the
cartridge.
Data backup issues
EMDB-01 -
The USB drive version is not correct, or the USB drive is
not properly connected to the FREND System. In addition, the USB
drive or USB H port has somehow been damaged.
• Check to be sure the USB drive is
the version 1.1 required for the FREND
TM
System and that the USB
drive is connected properly to the USB H port. If after trying a second
time with the same USB drive the error message is repeated, try a
new USB drive version 1.1. If the problem persists, contact customer
support for assistance.