38
The FREND™ System checks
To assure reliable and accurate test results, please check the following:
Observe the external appearance and screen of the FREND™ Sys-
tem every time you use the device.
Check the signal input part of the FREND™ System once a month
regardless of the device use, using the Q.C. Cartridge and the Q.C.
code chip to specifically test the optics.
If there are any problems found in the results after the checkup, stop
using the device immediately and contact the supplier’s customer
support.
The FREND™ System maintenance supplies
The FREND™ System checklist
1. Laser power
2. Laser alignment
3. Calculate ratio
FREND
TM
System
maintenance
Maintenance
•
•
•
Q.C. Pouch
Q.C. Cartridge
Q.C. Code chip
System
Q.C. Pouch
[Components]
1 x Q.C. Cartridge 1 x Q.C. Code chip
Summary of Contents for Frend system