Test result issues (cont.)
EMTF-07 -
The lateral flow of the sample within the cartridge is
incomplete.
• Load the sample into a new cartridge and restart the test. If the
same error message is displayed, centrifuge the sample to remove
particulate matter and fibrin clots and retry with a new cartridge. If the
problem persists, try a different sample. If the problem still continues,
contact customer support.
Update issues
EMUF-01 -
There is no executable file (FREND.exe) on the USB drive.
• Save the executable file, FREND.exe, transfer it to the USB drive and then
retry the update process.
Trouble-
shooting
Maintenance