Test result issues (cont.)
EMTF-05 -
The lateral flow of the sample within the cartridge is
incomplete.
•
Load the sample into a new cartridge and restart the test. If the same error
message is displayed, centrifuge the sample to remove particulate matter and
fibrin clots and retry with a new cartridge. If the problem persists, try a different
sample. If the problem still continues, contact customer support.
EMTF-06 -
The lateral flow of the sample within the cartridge is
incomplete.
•
Load the sample into a new cartridge and restart the test. If the same error
message is displayed, centrifuge the sample to remove particulate matter and
fibrin clots and retry with a new cartridge. If the problem persists, try a different
sample. If the problem still continues, contact customer support.