12. Prepare the cartridge and patient samples.
• For patient samples, collect 2-4 mL of blood in a 5 mL tube. Spin the tube in a
centrifuge and collect the serum or plasma.
13. After entering the patient information or ID on the sticker of the
cartridge, inject into the inlet of the cartridge, using a calibrated micro-
pipette and disposable tip, the volume of serum or plasma specified in
the package insert sheet for the analyte being measured.
14. Press the
TEST
button on the main screen and the system
moves to the ‘Patient ID’ screen automatically.
FREND
TM
System
operating
procedure
Operation
Caution:
Please be careful not to overfill.