46
Cartridge insertion issues
EMCI-04 -
The direction of the inserted cartridge is not correct.
• Re-insert the cartridge correctly. After insertion, check to make sure the
cartridge is properly and completely inserted.
Barcode issues
EMBR-01 -
The barcode printed on the cartridge may have been
damaged so the reader is unable to interpret the code correctly.
• Remove the damaged cartridge from the system. Load a new cartridge with
sample and reinsert the cartridge into the cartridge slot. Discard the damaged
cartridge.
EMBR-02 -
The barcode information related to the cartridge lot number
printed on the cartridge is not readable by the system reader.
• Manually enter the lot number of the cartridge using the keypad. The lot
number is printed directly above the barcode on the cartridge.