Test result issues
EMTF-01 -
The lot number of the inserted cartridge is different from
the lot number of the installed code chip.
• Check that the lot number printed on the cartridge is the same as
the lot number on the code chip. If not, restart using a code chip and
cartridge of the same lot number. If the original set were the same
lot number, try a new code chip and cartridge pair in case either was
damaged.
EMTF-02 -
Unable to output a test result because the system detected
readings in the cartridge during analysis that were abnormal so a
result could not be calculated.
• Load the same sample into a new cartridge and load the cartridge into
the system. If the problem is repeated, try a sample from a different source
(different patient). If the problem continues, contact customer support.