67
• An index between -0.5 and +0.5 indicates balanced water
• An index of more than +0.5 indicates scale-forming water
• An index below -0.5 indicates corrosive water
You can use the calculator to adjust any of the water balance
parameter variables used in the calculation to analyze the effect of the
change on water balance.
The LSI calculator function generates an accurate saturation index
value that indicates the state of the system at the time the calculation
is made.
C. Hardness Units
NOTE: The 9P MUST be in the HARDNESS FUNCTION for hardness
unit preference selection.
The hardness unit preference is set simultaneously for both functions
that capture a hardness value. That means when you set the unit
preference in Hardness Function, it is set for the LSI Calculator
function at the same time.
The LSI Calculator allows you to calculate with either ppm or grains as
hardness units.
The hardness unit conversion is based on the following equivalency:
17.1 ppm (mg/L) = 1 grain
XXV. SOFTWARE VERSION
Contact the Myron L
®
Company to see if a software upgrade is available.
1.
Press any parameter key.
2.
Press
key until three numbers are displayed as shown
in Figure 35.
3.
Press any parameter key; instrument will time out in 15