24
1.
Press
,
or
to select the parameter on
which you wish to change the solution type.
2.
Press and hold
key for 3 seconds to make “SEL”
appear (see Figure 1). For demonstration purposes, all 4
solution types are shown simultaneously.
3.
Use the
or
key to select type of solution desired
(ref. Solution Characteristics, pg. 59). The selected solution
type will be displayed: KCl, NaCl, 442
TM
or User.
4.
Press
to accept new solution type.
E. Application of User Solution Type
1.
User Programmable Temperature Compensation
(Tempco)
This feature allows you to change your Ultrameter
III
’s temperature
compensation factor to another factor between 0-9.99%/°C (ref.
Temperature Compensation, pg. 56). This feature does not apply to pH
or ORP.
a.
As in Procedure to Select a Solution, above, select User mode.
b.
With User mode now selected, press
. You may now
adjust a temperature compensation from .00%/°C to 9.99%/°C,
by pressing
or
. See example in Figure 2, pg. 25.