B. Setting User mode Calibration “Linc”
The Linc function sets or “links” the calibration gain factor of a Standard
Solution to the User solution mode. Once set, the “Linc” will stay intact
with future calibrations unless the Linc has been canceled. For more
information on canceling the User mode Calibration Linc refer to the
section C. Canceling User mode Calibration “Linc”, pg. 46.
Follow the steps below to set either the KCl, NaCl or 442
TM
calibration
factor to the User solution mode.
1.
Press measurement key desired to be “Linked”, i.e.,
,
or
.
2.
Place the Ultrameter
III
in User mode (ref. SOLUTION
SELECTION, pg. 23, for selecting the User mode).
3.
Press
arrow key until the menu “Linc” appears (see
Figure 27).
4.
Press
key. The instrument will display “SEL” and the
“User” Icon (see Figure 28).
Any additional display of KCl, NaCl or 442
TM
icons indicates
a “Linc” between the User solution and the other solution
displayed.