13
over the conductivity cell being careful
not to touch the tip to the existing
solution. Depress the top of the pipette
to the SECOND STOP (all the way
down) being careful to keep the tip over
the cell. Release the plunger button and
let it return to the REST position.
2.
T-Plunger Instructions
To PLUNGE the cell:
Clean and dry the T-plunger. With the
cell extender installed and solution in
the cell, insert the tip of the T-plunger
in the cell extender until the arms of the
T-plunger are flush against the rim of the
cell extender. Solution will overflow the
cell.
Keeping the arms of the T-plunger flush
against the rim of the cell extender, rotate
the T-plunger from side to side, allowing
the arms of the T-plunger to slide along
the rim of the cell extender.
Remove the T-plunger from the cell by
lifting it straight up out of the cell, being
careful to keep it centered over the mouth
of the cell extender. When the T-plunger
is completely out of solution, tap the
T-plunger on the inside rim of the cell
extender to dispel any clinging solution
back into the cell.
Remove the T-plunger completely from
the cell area.
3.
Mixing Solution in Cell
a. Agitate
When the display says to agitate (“AGit”):
Grasp the instrument on both sides of
the keypad (keypad facing up) with your
hands so you don’t accidentally drop or
release it. Agitate the solution in the cell
by swinging the cell cup area of the 9P up
and down five or more times.
b. Hold
When the display says to hold (“HOLd”):
Hold the instrument steady in the
horizontal position while the instrument
records a measurement.
Insert the T-plunger all the way
down into the cell extender then
rotate it from side to side.
Hold the T-plunger over the cell
then tap firmly so that solution is
dispelled back into the cell cup.
“AGit”: When “AGit” displays, swing
the cell end of the 9P up and down
five or more times.
“HOLd”: When “HOLd” displays, hold
the 9P still with bottom of case parallel
to the ground.